High Resolution Careray 17"17 Digital Wireless X-ray Flat Panel Detector for Medical Imaging
MSLCV35 is 17*17 movable flat panel detector. This manual describes the parameter specification, installation way, storage and transportation, precautions and other aspects of the flat panel detector.
Main Features •Advanced encapsulation and internal and external frame design allow the detector to be used in a wide temperature range and vibration environments; •Pixel size: 140μm, image array: 3072 x 3072, 9.43 million pixels; •High sensitivity, low X-ray dose; fast exposure (less than 5 ms); •Excellent dynamic range, 9×107 saturated electrons; maximum linear exposure dose: more than 100μGy; •Multi-stage gain adjustable; it is easy for users to get the best digital images.
Main Advantages •Completely independent development, cost-effective; •The most advanced flat panel detector technology with cesium iodide directly coupled with amorphous silicon can be used; •TFT module from an international famous manufacturer is selected; •Timely, effective and comprehensive quality services; •It can be widely used in the fields such as medical diagnostic imaging system, animal diagnosis, industrial NDT and safety inspection.
Main Structure The amorphous silicon flat panel detector is mainly composed of an X-ray scintillation screen with high absorption rate, a large-size amorphous silicon photodiode / TFT flat panel, a detector electronic control system, etc.







Imaging Display Details


Specification
| Scintillator Direct Deposit Csl | TI |
| Active Area | 430mm X 430 mm |
| Image Matrix Size | 3072 X 3072 |
| Pixel Pitch | 140 um |
| A/D Conversion | 16 bits |
| Data Interface | Gigabit Ethernet/WiFi (2.4 /5GHz) |
| Acquisition Time | 2S-4S |
| Weight | 4.2 kg |
| Size | 460 mmx460 mmx15 mm |
| Power Consumption Max | 64W |
| Power Supply | 100-240V |
| Frequency | 50/60Hz |
Installation Instructions
1、The flat panel detector must be installed by specialized personnel. 2、When installing the detector system, prevent the detector from slipping off. 3、During the installation of detector, reduce or avoid detector vibration and impact as much as possible!
Installation Steps for Detector X-ray flat panel detector must be installed in the film magazine before use. Specific installation steps are as follows: •Put the detector into the film magazine, and fasten the detector with pressing pieces around the film magazine; •Lead the detector connecting wire out of the film magazine through the film magazine wire hole; •Connect the external connecting wire in the air and connect it to the power supply box; •Connect the power supply box with DROC workstation through the network cable; •Connect the power supply box with the high voltage generator through a hard synchronous line; •Plug in the power cord and AC cable and turn on the detector, and the detector can work normally.
Introduction to DR System Connection The whole DR system mainly includes X-ray ball tubes, an X-ray generator, a flat panel detector and a DROC workstation, and the connection among the flat panel detector, the high voltage generator (Ruidi high voltage generator) and DROC workstation is shown in Figure 7 below. The detector is directly connected to the power supply box, and the gigabit Ethernet cable and the synchronous exposure cable transferred by the power supply box are connected to DROC workstation and the high voltage generator respectively. 220V power of the power supply box can be supplied from the grid power on the wall or the rack.

Package and Delivery

| Product Name | Flat Panel Detector |
| Place of Origin | Guangdong, China |
| Warranty | 1 Year |
| Power Source | Electric |
| After-sale Service | Online Technical Support |
| Shape | Rectangular |
| Color | Black |
| Scintillator | Direct Deposit CsI TI |
| Pixel Pitch | 140 μm |
| Active Area | 430mm X 430 mm |
| Image Matrix Size | 3072 X 3072 |
| A/D Conversion | 16 bits |
| Data Interface | Gigabit Ethernet/WiFi (2.4/5GHz) |
| Acquisition Time | 2S-4S |
| Weight | 4.2 kg |
| Dimensions | 460 mm x 460 mm x 15 mm |
| Maximum Power Consumption | 64W |
| Power Supply | 100-240V |
| Frequency | 50/60Hz |
| Shelf Life | 2 Years |
| Quality Certification | CE |
| Instrument Classification | Class II |
| Application | X-ray photography |